Evaluation of Patient-Specific Quality Assurance (PSQA) using Octavius 4D in IMRT Planning for Nasopharyngeal Cancer Cases

Rizki Budi Rahayu, Rustika Alifiani, Aditya Prayugo Hariyanto, Levina Almira, Endarko Endarko

Abstract

Patient-Specific Quality Assurance (PSQA) is an important procedure for verifying the accuracy of dose planning before administering radiation therapy to patients. This study investigated the effects of progressively stricter OAR constraints on fluence complexity and Patient-Specific Quality Assurance (PSQA) performance in nasopharyngeal cancer IMRT as a preliminary case-based investigation. Three IMRT plan variations were created for two patients (tumor volumes: 56.16 cm³ and 75.35 cm³) using TPS Monaco: Plan0 (no OAR constraints), Plan1 (moderate constraints), and Plan2 (strict constraints) for the parotid glands, brainstem, and spinal cord. PSQA was performed using the PTW Octavius 4D phantom with gamma index criteria of 3%/3 mm, 3%/2 mm, and 2%/2 mm, in accordance with the AAPM TG-218. All plans achieved GPR values above the 95% tolerance limit for the 3%/2 mm criterion (95.5-99.8%) and above 90% for the 2%/2 mm criterion, confirming clinically acceptable deliverability across all optimization levels. Notably, the GPR in the patient with a larger tumor volume (ID2, 75.35 cm³) did not decrease with stricter constraints. The Plan2 GPR (99.5%) exceeded the Plan1 GPR (98.5%), whereas the patient with a smaller tumor volume (ID1, 56.16 cm³), whose GPR declined consistently from Plan0 to Plan2. This divergence reveals that the relationship between OAR constraint stringency and plan deliverability is patient-specific and governed by the geometric interaction between the tumor volume and surrounding OARs; larger targets afford the TPS greater dosimetric flexibility, thereby reducing the need for narrow MLC segments even under strict constraints. These findings suggest that Octavius 4D-based PSQA serves not only as a routine delivery verification tool but also as a patient-specific plan selection tool, and that the 3%/2 mm criterion is recommended as the primary clinical evaluation parameter for nasopharyngeal IMRT.

Keywords

Gamma Index; IMRT; Nasopharyngeal Cancer; Octavius 4D; PSQA

Full Text:

PDF

References

[1] Fischer-Valuck B. W., Rao, Y. J., and Michalski JM. 2018. Intensity-modulated Radiotherapy for Prostate Cancer. Transl. Androl. Urol. 2018;7(3):297-307, doi: 10.21037/tau.2017.12.16

[2] Pacelli, R., Caroprese, M., Palma, G., Oliviero, C., Clemente, S., Cella, L., and Conson, M. 2019. Technological Evolution of Radiation Treatment: Implications For Clinical Applications. Seminars in Oncology Volume 46, Issue 3, June 2019, Pages 193-201

[3] Yoganathan, S. A., Mani, K. R., Das, K. J. M., Agarwal, A., and Kumar, S. 2011. Dosimetric Effect of Multileaf Collimator Leaf Width in Intensity-Modulated Radiotherapy Delivery Techniques for Small- And Large-Volume Targets. J Med Phys. 2011 Apr-Jun;36(2):72–77, doi: 10.4103/0971-6203.79690

[4] Zhu, H., Zhu, Q., Wang, Z., Yang, B., Zhang, W., and Qiu, J. 2022. Patient-Specific Quality Assurance Prediction Models Based on Machine Learning for Novel Dual-Layered MLC Linac. Medical Physics Volume 50, Issue 2 pp. 1205-1214. 2022, https://doi.org/10.1002/mp.16091

[5] Ni, Q., Chen, L., Tan, J., Pang, J., Luo, L., Zhu, J., and Yang, X. 2024. Predicting The PSQA Results of Volumetric Modulated Arc Therapy Based on Dosiomics Features: A Multi-Center Study. Front. Phys. 12:1387608, doi: 10.3389/fphy.2024.1387608

[6] Listiono, F. I., Noor, J. A. E., Herwiningsih, S., Dwiyono, A. G., Martono, S., Nugroho, R. S., and Petrarizky, A. J. 2025. Implementation of AAPM TG-218 for Patient Specific Quality Assurance (PSQA) in The Case of Thoracic Target Region Using IMRT Technique with EPID ASi-1200“, Indonesian Physical Review, vol. 8, no. 3, p 690 699, 2025, https://doi.org/10.29303/ip r.v8i3.502.

[7] Das, S., Kharade, V., Pandey V, et al. 2022. Gamma Index Analysis as a Patient-Specific Quality Assurance Tool for High-Precision Radiotherapy: A Clinical Perspective of Single Institute Experience. Cureus 14(10): e30885, doi: 10.7759/cureus.30885

[8] Mrčela, I., Gregov, M., Matanić, A., Budanec, M., Murgić, J., Jakšić, B., Prpić, M., Sečan, A. P., and Frobe, A. 2022. Dosimetric Verification of Intensity Modulated Radiotherapy (IMRT) Treatment Plans for Prostate Cancer Patients. Acta Clin Croat (Suppl. 3) 2022; 61:21-27, doi: 10.20471/acc.2022.61.s3.3

[9] Low, D. A., Harms, W. B., Mutic, S., and Purdy, J. A. 1998. A Technique for The Quantitative Evaluation of Dose Distribution. Med Phys. 1998, 25:656-61, doi: 10.1118/1.598248

[10] Teichert, K. et al. 2019. Targeted Multi-Criteria Optimisation in IMRT Planning Supplemented by Knowledge Based Model Creation. Oper. Res. Heal. Care, vol. 23, p. 100185, 2019.

[11] Shin, H. B., Noh, Y. Y., Kim, H., Kim, C., Sung, K. H., and Kim, D. W. 2024. Correction Method for Depth and Field Size Dependence in Octavius 1000 SRS Patient-Specific QA. Radiation Physics and Chemistry, 223 (2024) 111892.

[12] Shen, Z., Tan, X., Li, S., Tian, X., Luo, H., Wang, Y., and Jin, F. 2021. Correlation between the γ passing rates of IMRT plans and the volumes of air cavities and bony structures in head and neck cancer. Radiation Oncology, 16(1):148, doi: 10.1186/s13014-021-01861-y

[13] Bin, S., Zhang, J., Shen, L., Zhang, J., and Wang, Q. 2023. Study of the prediction of gamma passing rate in dosimetric verification of intensity-modulated radiotherapy using machine learning models based on plan complexity. Frontiers in Oncology, 13:1094927, doi: 10.3389/fonc.2023.1094927

[14] Chiavassa, S., Bessieres, I., Edouard, M., Mathot, M., and Moignier, A. 2019. Complexity metrics for IMRT and VMAT plans: a review of current literature and applications. British Journal of Radiology, 92(1102):20190270, doi: 10.1259/bjr.20190270

[15] Dogan, N., Mijnheer, B.J., Padgett, K., Nalichowski, A., Wu, C., Nyflot, M. J., Olch, A. J., Papanikolaou, N., Shi, J., Holmes, S. M., Moran, J., and Greer, P. B. 2023. AAPM Task Group Report 307: Use of EPIDs for Patient-Specific IMRT and VMAT QA. Med Phys. 2023;50:e865–e903., doi:10.1002/mp.16536

[16] Deng J., Liu S. Y., Huang Y., Li X., and Wu X. 2024. Evaluating AAPM-TG-218 Recommendations: Gamma Index Tolerance and Action Limits in IMRT and VMAT Quality Assurance Using SunCHECK. J. Appl. Clin. Med. Phys., vol. 25, no. 6, pp. 1–12, 2024.

[17] Ezzell, G. A., Burmeister J. W., Dogan, N., LoSasso, T. J., Mechalakos, J. G., Mihailidis, D., Molineu, A., Palta, J. R., Ramsey, C. R., Salter, B. J., Shi, J., Xia, P., Yue, N. J., and Xiao, Y. 2009. IMRT Commissioning: Multiple Institution Planning and Dosimetry Comparisons, A Report From AAPM Task Group 119. Med. Phys. 36 (11) November 2009.

Refbacks

  • There are currently no refbacks.